Rapid Thermal Cycling Chamber
Product Usage :Rapid Thermal Cycling Chamber (Rapid Ramp Change Temperature Test Chamber) give the maximum durability and performance for your most demanding test requirements.It is the necessary test equipment for aviation, Aerospace, automotive, electronics, military and solar panel industry etc.It can be used to test and determine the parameters and performance of electrical, electronic products and materials in the most extreme temperature and humidity conditions.
Apply to:Cyclic Temperature Rapid Change Test
Customizable :Specific product specifications and parameters can be customized according to the customer's actual situation or customization. You can contact us at any time to obtain quotations and solutions.
Parameter Details
Sizes of Rapid Ramp Change Thermal Chambers | |||||
Model
|
Volume (L)
|
Temperature Range
(°C) |
Rated Power
|
Internal Dimension (mm) D*W*H
|
External Dimension (mm) D*W*H
|
LRHS-101B(R5)-LJ
|
101
|
A: -20~150°C
B: -40~150°C C: -60~150°C D: -70~150°C E: or customized |
12.0kW
|
450*450*500
|
1160*1000*1610
|
LRHS-225B(R5)-LJ
|
225
|
14.0kW
|
500*600*750
|
1210*1150*1870
|
|
LRHS-504B(R5)-LJ
|
504
|
16.0kW
|
700*800*900
|
1260*1340*2070
|
|
LRHS-800B(R5)-LJ
|
800
|
18.0kW
|
800*1000*1000
|
1370*1550*2170
|
|
LRHS-1000B(R5)-LJ
|
1000
|
20.0kW
|
1000*1000*1000
|
1540*1490*2160
|
|
Note: In the Model, the number means volume of Rapid Thermal Cycling Chamber working room, and ABCD means the temperature range. For example, model LRHS-101B(R5)-LJ means the volume of this Rapid Thermal Cycling Chamber is 101L, and its temperature range is -20°C~+150°C. LRHS-225B(R5)-LJ means the volume is 225L and the temeprature range is -60°C~+150°C |
|||||
Basic Specification | |||||
Performance
|
Temperature Range | A: -20~150°C, B: -40~150°C, C: -60~150°C, D: -70~150°C, E : or customized | |||
Temperature Fluctuation | ±0.5°C | ||||
Temperature Uniformity | ±2.0°C | ||||
Ramp Rate | ﹣55°C~85°C ≤15°C/min for heating (non-linear, based on on-load) 85°C~ -55°C≤15°C /min for cooling (non-linear, based on on-load) Note: can be customized as per requirement |
||||
Material
|
External Casing | Cold-rolled A3 (Q235) steel plates with plastics painted. | |||
Interior Working Room | Stainless Steel Plates SUS304 with miror surface | ||||
Heat Isolation | Polyurethane foams and fine glass fibres | ||||
Illumination Lamp | Philips® one in per standard test chamber. | ||||
Casters | 4pcs, with PU covered for protection. Height adjustable and direction universal. | ||||
Viewing Window | Multi-layer hollow toughened glass with conductive films for heating to defrost | ||||
Door | Manually operated sliding door with unlock button | ||||
Heating & Cooling System
|
Heater | Steipped Nickel-chromium Wire Heater | |||
Air Blower | Anti-temperature stainless fan blades with low noise | ||||
Compressor | Hermetically-sealed Techmesh® or semi-sealed Bitzer® or GEA® or equivalent compressors | ||||
Cooling Method | Single or dual cooling (Air-cooled or Water-cooled) | ||||
Refrigerant | R404A or/and R23 (eco-friendly) | ||||
Condenser | Plate-type heat exchanger | ||||
Control System
|
Controller | 7" Programmable colorful touch-screen display LCD English Display Microcomputer integrated controller |
|||
Functions | Data Record (Max 600 days x 24hrs), History Curve, USB data download & upload, Remote controlling through PC, Automatic protection alarms, etc. | ||||
Program | 1~1000 programs, 1~999 steps for per program, Max 99hrs 59mins for per program. 10 programs can be interlinked. | ||||
Temperature Sensor | Platinum Resistance.PT100Ω | ||||
Running Model | Constant/Fix-value Model and Programmable Model | ||||
I/O Interface | Communicate port (RS485) to connect test chamber to a PC | ||||
Standard Configuration | Test Chamber x 1, Specimen Holder Shelf x 2, φ100mm Cable Access Port x 1 with 1 soft plug, EMO x 1, Tri-color Indicator x 1, etc. | ||||
Safety Protection | Power leakage protection/Compressor over-heat, over-load or over-current protection/Dry burning protection. | ||||
Voltage | 380V/440V±10%, 50Hz/60Hz or as per requirement | ||||
Ambient Enviromental for application | 5°C~+30°C ≤ 85%R.H | ||||
Test Standards (not limited to) | IEC60068-2-1
IEC60068-2-2
IEC60068-2-78
IEC60068-2-30
|
Feature
1. Programmable Touch Screen Controller
2. Custom Sizes Available
3.Running Time Meter
4. Wide Temperature and Humidity Ranges
5. Capable of Handling Extreme Temperature and Humidity
6. High Performance Even Under Load
7. Incandescent Interior Lighting
8. Modular,Robust Reliable Construction
9. Stainless Steel Interior
10. Painted Galvanized Steel Exterior For Long Life
11. Heavy-duty Flooring to Support Vehicles
12. Mechanical Cooling
13. Comprehensive Safety Features 14.Refrigeration Gauges
15. Refrigeration Sound Deadening Package
16. Low or Zero Ozone Depletion Refrigerants
17. All Circuits Fused or Circuit Breaker Protected.
18. All Wiring Numbered or Color Coded
19. Compressor Overload Protection
20. Prefabricated for On-Site Installation
Test standard
IEC 60068-3-12:2022
IEC 60068-2-38:2021
IEC 60068-3-5:2018
MIL-STD-810H:2019 Method 507.6
ISO 9022-3:2022
ISO 13007:2023
IPC-9701C:2021
UL 61010-1:2019
EN 14597:2022
EN IEC 61010-1:2021
American Standards:
UL 61010-2-012:2023
UL 61010-1:2019
Chinese Standards:
SJT 31129-1994 Requirements and Inspection and Evaluation Methods for the Integrity of High and Low Temperature Testing Equipment
JB-T 8250.5-1999 Test Methods for High and Low Temperature Tests of Cameras
GJB 4.6-1983 Environmental Tests for Shipboard Electronic Equipment: Alternating Humidity and Heat Test
GB-T 5170.5-2008 Inspection Methods for Environmental Testing Equipment for Electrical and Electronic Products: Humidity and Heat Test
GBT 11158-1989 Technical Conditions for High-Temperature Test Chambers
Submit your requirements
Application
Scope of application-
New Energy
Test chambers have extensive applications in the n… -
Rubber products
Before rubber products leave the factory, it is es… -
Electronic and electrical products
The purpose of environmental testing for electrica… -
Automotive Industry
The automotive industry must conduct numerous envi… -
Aerospace and Defense Industry
As aerospace technology advances and the defense i…